Connect to the host power supply, and the green main power indicator on the display lights up.
Connect high-pressure air and oxygen.
The ventilator system switch is located on the back of the host, turn on the switch.
After the ventilator starts and completes the self-test, the system enters standby mode and displays the select patient menu on the display.
When the ventilator is in standby mode, the patient settings menu will be displayed on the normal screen.
After each verification is completed, the result will be displayed next to it.All verification is complete.
After that, ‘Verification completed’ will be displayed, and the highlighted area will be moved to the delete trend.
The ventilation mode can be changed during standby or during the operation of the ventilator.
Note that the atomizer cable should not be soaked or autoclaved.
Clean All outer surfaces with a damp cloth stained with a mild detergent.
Please clean the main unit and the filter on the screen regularly. For the rest of the daily maintenance, please refer to the “Maintenance” instructions.
Note: The exhalation flow sensor, the exhalation valve housing (including diaphragm, O-ring, and water trap), and the Aeroneb Pro atomizer can all be sterilized at 134°C high pressure.
It is recommended that the exhalation flow sensor be soaked in 75% alcohol for disinfection, and should not be rinsed against the faucet.