How to operate the ventilator of Maquet servo-i and servo-s?

hamilton ventilator parts
Common faults and maintenance methods of PB840 ventilator
11月 4, 2022
o2 sensor medical
GE Engstrom Caerstation Ventilator operation method
11月 6, 2022
Show all
maquet ventilator parts

maquet ventilator parts

1. Turn on the ventilator power:

  •  Connected to power supply: host, air compressor, humidifier;
  • Connected to the air source: high-pressure oxygen, high-pressure air;
  • Connect the sterilized exhalation box, breathing tube, humidification tank, and add an appropriate amount of sterile distilled water;
  • Turn on the host power supply and the air compressor power supply (except for no air compressor);
  • Press the interface “Check before use” to make a self-test, (pay attention to the use of the original test pipeline) and make sure that all can pass correctly.;
  • For child mode, please set the pipeline compliance compensation to (no);
  • Select the ventilation mode and adjust the breathing parameters;
  • Press the ventilation/stop ventilation button to start ventilation, and use the simulated lungs for ventilation for about 1 minute to ensure that the parameter settings are correct.;
  • Connect the patient for normal ventilation;
  • Ventilate for about 30s, and after the ventilation is stable, set the appropriate upper and lower alarm limits.

2.Shut down ventilator:

  • Press the ventilation/stop ventilation button to stop ventilation;
  • Disconnect the power supply and air source;
  • Remove the exhalation box, breathing tube, humidification tank, humidifier accessories, etc., and use suitable methods for sterilization and disinfection;
  • Clean the ventilator;
  • Store the ventilator in a dry, room temperature room.

3. Cleaning process:

 Respiratory circuit disinfection process

  •  Discard disposable parts: disposable pipes, filters, etc.;
  • Detoxifying parts: silicone pipeline, humidification tank, exhalation box, etc.;
  • Rinse all parts with hot water below 35 degrees;
  • Soak and disinfect with disinfectant;
  • Disinfectants can be used: alcohol, phthalaldehyde-containing solution (CidexOPA), Haisani instrument cleaning disinfectant, DD1 multi-enzyme sterilization cleaner, Yaniosede 1000 disinfectant;
  • Clean the parts in sterile water to remove all residues;
  • Drying: It can be dried naturally for more than 24 hours, or it can be dried in a 70-degree drying oven for 1 hour.

Respiratory circuit sterilization process (not recommended)

  • Discard disposable parts: disposable pipes, humidifying paper, filters, etc.;
  • Sterilizable parts: silicone pipeline, humidification tank, exhalation box, etc.;
  • Rinse all parts with hot water below 35 degrees;
  • Autoclave: peak temperature of 134 degrees, 4 minutes;
  • Incorrect autoclaving may cause damage to the exhalation box.Please use high-quality sterilization equipment and professional personnel to operate.

Daily maintenance and maintenance of equipment

  • If the use time reaches 7 days, please check the air compressor and perform maintenance (except for no air compressor);
  • Clean the dust filter at the bottom and rear of the air compressor (except for no air compressor);
  • Exclude stagnant water from the stagnant water bottle behind the air compressor (except for no air compressor)
  • The surface of the equipment can be cleaned with a damp cloth;
  • If the use time reaches 5000 hours, please maintain the host of the equipment.